SITRANS P, MS series
115
C79000-B5674-C40-04
Technical Data
■
Technical data
(continued)
Certificates and approvals
Classification according to pressure
equipment directive
(DGRL 97/23/EC):
For gases of fluid group 1 and
liquids of fluid group 1;
complies with requirements of
article 3, paragraph 3 (sound
engineering practice)
Explosion protection
• Intrinsically-safe version
- Intrinsic safety "i"
PTB 99 ATEX 2122
- Identification
II 1/2 G EEx ia/ib IIC/IIB
T4/T5/T6
- Permissible ambient temperature
-40 to +185 °F/+158 °F/+140°F
- Connection to certified intrinsi-
cally-safe circuits with maximum
values:
U
i
= 30 V, I
i
= 100 mA
P
i
= 750 mW, R
i
= 300
Ω
- Effective internal inductance
L
i
=
1 mH
- Effective internal capacitance
C
i
=
6 nF
• Flameproof design
- Flameproof enclosure "d"
PTB 99 ATEX 1160
- Identification
II 1/2 G EEx d IIC T4/T6
- Permissible ambient temperature
-40 °F to +185 °F/+140°F
- Connection to circuits with values: U
H
= DC 10.5 to 45 V
• Zone 2 version
- Type of protection "n"
TÜV 01 ATEX 1696 X
- Identification
II 3 G EEx nA L IIC T4/T5/T6
- Permissible ambient temperature
-40 to +185 °F/+158 °F/+140°F
- Connection to circuits with values: U
H
= DC 10.5 to 45 V
• Explosion protection to FM
Certificate of Compliance
3008490
- Identification (XP/DIP) or (IS); (NI) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; CL
I, ZN 0/1 AEx ia IIC T4...T6; CL I,
DIV 2, GP ABCD T4...T6; CL II,
DIV 2, GP FG; CL III
- Permissible ambient temperature
-40 to +185 °F/-40 to +158 °F/
-40 to +140°F
- Entity parameters
Acc. to control drawing
A5E00072770A:
U
i
= 30 V,
I
i
= 100 mA,
P
i
= 750 mW,
R
i
= 300
Ω,
L
i
=
0,4 mH,
C
i
=
6 nF
• Explosion protection to CSA
Certificate of Compliance
2000.1153651
- Identification (XP/DIP) or (IS)
CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex
ia IIC T4...T6; CL I, DIV 2, GP
ABCD T4...T6; CL II, DIV 2, GP
FG; CL III
- Permissible ambient
temperature
-40 to +185 °F/-40 to +158 °F/
-40 to +140°F
- Entity parameters
Acc. to control drawing
A5E00072770A:
U
i
= 30 V,
I
i
= 100 mA,
P
i
= 750 mW,
R
i
= 300
Ω,
L
i
=
0.4 mH,
C
i
=
6 nF
Communication
Load when connecting a
• HART communicator
230 to 1100
Ω
• HART modem
230 to 500
Ω
Cable
2-wire screened:
≤
3.0 km (1.86 miles)
Multi-core screened:
≤
1.5 km
(0.93 miles)
Protocol
HART, version 5.x
PC/laptop requirements
IBM-compatible,
main memory > 32 Mbyte,
hard disk > 70 Mbyte,
RS 232 interface,
VGA graphics
Software for PC/Laptop
WINDOWS 95/98/NT 4.0
and SIMATIC PDM
Summary of Contents for MS Series
Page 2: ......
Page 4: ...SITRANS P Serie MS 4 C79000 B5674 C40 04 ...
Page 10: ...SITRANS P Serie MS 10 C79000 B5674 C40 04 ...
Page 16: ...SITRANS P Serie MS 16 C79000 B5674 C40 04 Technische Beschreibung ...
Page 50: ...SITRANS P Serie MS 50 C79000 B5674 C40 04 Inbetriebnahme ...
Page 51: ...SITRANS P Serie MS 51 C79000 B5674 C40 04 Technische Daten 7 Siehe folgende Seiten ...
Page 56: ...SITRANS P Serie MS 56 C79000 B5674 C40 04 Pflege und Wartung ...
Page 60: ...SITRANS P Serie MS 60 C79000 B5674 C40 04 Zertifikate ...
Page 66: ...SITRANS P Serie MS 66 C79000 B5674 C40 04 Index ...
Page 72: ...SITRANS P MS series 72 C79000 B5674 C40 04 ...
Page 78: ...SITRANS P MS series 78 C79000 B5674 C40 04 Technical Description ...
Page 112: ...SITRANS P MS series 112 C79000 B5674 C40 04 Commissioning ...
Page 113: ...SITRANS P MS series 113 C79000 B5674 C40 04 Technical Data 7 See following pages ...
Page 118: ...SITRANS P MS series 118 C79000 B5674 C40 04 Maintenance and Care ...
Page 122: ...SITRANS P MS series 122 C79000 B5674 C40 04 Certificates ...
Page 128: ...SITRANS P MS series 128 C79000 B5674 C40 04 Index ...
Page 129: ......