2 - 1
Siemens AG
SPL1-130.840.02
Page 1 of 14
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
Info, E00 and E99
2
Water system, E10
–
E12
2
LITHOSTAR MODULARIS circuit diagram J1048, sheets 13 and 14.
Error messages from the SIREMOBIL Iso-C have not been listed.
(Refer to the SIREMOBIL Iso-C service documents)
E00
Error Text
:
System was switched on.
Possible
cause/action:
E99
Error Text
:
The system was switched off without deleting the therapy data.
Possible
cause/action:
Press the Reset button.
E10
Error Text
:
Cooling system losing water.
Possible
cause/action:
Is water running out of the hoses?
P
min
-switch o.k.?
Board D3.X3 connection o.k.?
Fill the cooling circuit.
E11
Error Text
:
Maximum pressure in the cooling circuit exceeded.
Pressure switch P
max
responded.
Possible
cause/action:
Are the hoses blocked, kinked?
P
max
-switch o.k.?
Board D3.X3 connection o.k?
Is the cooling unit operating?
Check the fill level in the cooling circuit.
E12
Error Text
:
Minimum pressure in the cooling circuit exceeded.
Pressure switch P
min
responded.
Possible
cause/action:
Is water running out of the hoses and hose connections?
Air in the cooling system?
Check the fill level in the cooling circuit.
NOTE
LITHOSTAR MODULARIS Error List
2