LITHOSTAR MODULARIS
4 - 13
Siemens AG
SPL1-130.840.02
Page 13 of 34
MODULARIS Uro Plus
Medical Solutions
Rev. 03
04.05
CS PS 24
IWAKI cooling pump
4
Removal
4
•
Switch the system off.
•
Open the system cover.
•
Remove the fasteners on the pump.
•
Remove the motor connections on the terminal block.
•
Loosen the screws to remove the pump.
Installation
4
•
Tighten the pump.
•
Connect the motor connections.
•
Connect the fasteners on the pump (The black-marked water hoses must be connected).
•
Fill the cooling circuit (refer to the section "Filling the cooling circuit").