Parameters
12.13 Test and Diagnostic (P900 - P951)
BW500 and BW500/L
Operating Instructions, 12/2016, A5E33482052-AD
189
12.12.5
P795 SmartLinx Protocol
This parameter is used to identify the protocol when SmartLinx is used. If you are not using
SmartLinx, this parameter is not functional. Please see the associated SmartLinx instruction
manual for a full description of this parameter.
If a SmartLinx module is not configured, P795 will have a value of either 255 or 0.
12.12.6
P799 Communications Control
Assigns programming control through the local keypad (or Dolphin Plus, P770 = 1), or through a
Modbus protocol (P770 = 2 or 3)
Entry:
0 = local
1 = modbus
12.13
Test and Diagnostic (P900 - P951)
Note
These parameters are used for test and diagnostic purposes.
12.13.1
P900 Software Revision
Displays the EPROM (Flash ROM) software revision level.
12.13.2
P901 Memory Test
Tests the memory. Test is initiated by scrolling to the parameter or repeated by 'pressing enter'
Display:
PASS = normal
FAIL = consult Siemens.