Commissioning and operation
5.4 Starting the PC and operating software
MicroSAM
74
Operating Instructions, 06/2012, C79000-G5376-C560-07
5.4
Starting the PC and operating software
The factor setting for the IP address of the gas chromatograph is 161.218.51.xx. The subnet
mask is always set to 255.255.255.0.
Requirement
● Manually set the IP address of the PC to a value between 161.218.51.0 and
161.218.51.255 with the subnet mask 255.255.255.0.
The PC must not have the same IP address as the gas chromatograph.
We therefore recommend the IP address 161.218.51.254.
● The connection between the PC and gas chromatographs has been made, and the
Windows XP operating system has been started.
Procedure
Start the program "Maxum System Manager" (start bar Start ➜ Programs ➜ Maxum System
Tools ➜ Maxum System Manager).
Start the function "Add Unit" (Menu "System" ➜ Menu command "Add Unit") and enter the
new gas chromatographs.
Note
Operating software
The following are described in detail in the Maxum System Manager manual:
Installation of the Maxum System Manager and EZChrom operating software.
Establishment of connection between PC and gas chromatographs.
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...