Mounting
4.4 Mounting and connecting
MicroSAM
Operating Instructions, 06/2012, C79000-G5376-C560-07
65
4.4.11
Connecting of digital outputs and digital inputs
The following table is valid for devices from release version 4 onwards.
Core color Description
Usage
Comment
Circuit diagram
White/gray Digital input 1+
Sample flow,
24 V DC = OK,
0 V = too low
Gray/brown Digital input 2+
Application programmable
White/red
Digital input 3+ or
SSSI* SCL-Rx
Application programmable
Brown/red
Digital input 4+ or
SSSI* SDA-Rx
Application programmable
24 V DC to optocoupler
Second connection to
+24 V DC
9
Brown/blue Digital output 1+
Application programmable
White/blue Digital output 2+
Application programmable
White/pink Digital output 3+ or
SSSI* SCL-Tx
Application programmable
Pink/brown Digital output 4+ or
SSSI* SDA-Tx
Application programmable
Relay contact
0.5 A / 100 V / 10 W
Second connection to
+24 V DC
9
* The SSSI signals from devices are available with release version 10. The signals are used,
for example, to connect an IO extender (order no. A5E01494354). Refer to the
documentation of the IO extender for more on this. If SSSI is not used, the lines are available
for their original function as digital inputs and outputs.
The following table is valid for devices up to and including release version 3.
Core color
Description
Usage
Comment
White/gray
Digital output 1+
Gray/brown Digital output 1-
White/red
Digital output 2+
Brown/red
Digital output 2-
Relay contact
0.5 A / 100 V / 10 W
Brown/blue Digital input 1+
White/blue
Digital input 1-
Sample flow,
24 V DC = OK,
0 V = too low
White/pink
Digital input 2+
Pink/brown Digital input 2-
24 V DC to optocoupler
Arc-suppression diode with inductive loads
If you connect inductive loads (e.g. valves) for digital outputs, you must connect the arc-
suppression diodes in parallel.
Type of cable
Siemens AG recommends twisted-pair cables, 0.14 mm
2
(AWG 26).
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...