Mounting
4.4 Mounting and connecting
MicroSAM
64
Operating Instructions, 06/2012, C79000-G5376-C560-07
4.4.10
Fitting the power cable
The chromatograph is supplied with a 4-m long cable. The cable has a diameter of 18 mm.
Connect all cores of the power supply in accordance with the following table since one single
core is insufficient for the maximum current. The cable must be securely routed and
terminate in a terminal box with approved type of protection.
Appropriate cable gland: M32 with a terminal areas from 13 to 21 mm.
Core color
Description
White
+24 V DC
Brown
+24 V DC
Green
+24 V DC
Yellow
+24 V DC
Gray
GND 24 V DC
Pink
GND 24 V DC
Blue
GND 24 V DC
Red
GND 24 V DC
Black
Permissible cable length
The power supply on the chromatograph must be
● up to AS07: 24 V DC (-15 ... +10%)
● from AS08 onwards: 24 V DC (18.5 ... 30.2 V)
If the power supply delivers exactly 24 V DC, and if only one chromatograph is connected,
the following cable lengths are permissible between the power supply and the
chromatograph depending on the cable cross-section (distance = half the cable length):
Core cross-section
Distance
1.5 mm
2
40 m
2.5 mm
2
70 m
4.0 mm
2
110 m
6.0 mm
2
170 m
10.0 mm
2
280 m
16.0 mm
2
450 m
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...