Mounting
4.4 Mounting and connecting
MicroSAM
62
Operating Instructions, 06/2012, C79000-G5376-C560-07
4.4.8
Breathing device
The breathing device is used for pressure compensation during operation. Pressure
compensation is required due to temperature variations during operation
①
Breathing connector
Figure 4-3
Breathing device
Danger
DANGER
Danger of explosion through discharged gases
In the event of a fault, all gases supplied to the chromatograph (also samples and hydrogen
as carrier gas) can be discharged via the breathing connector or a connected line. In such a
case, there is a danger to personnel, equipment and the environment!
For this reason, take appropriate measures to prevent this.
Never connect the breathing connector to the exhaust gas system.
WARNING
Do not close off the breathing device
Ensure that the breathing device is not closed during operation, otherwise:
Pressure compensation is no longer carried out.
Ex protection is no longer guaranteed.
Housing protection IP65 is only guaranteed if the pipeline connected to
①
ends in a
room which contains neither splashing water nor dust.
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...