Mounting
4.4 Mounting and connecting
MicroSAM
58
Operating Instructions, 06/2012, C79000-G5376-C560-07
4.4.2
Gas lines
The gas lines are made of stainless steel with a 3 mm ∅ (1/8'').
● Use only very clean gas lines suitable for chromatography.
● Connect the gas lines to the gas chromatograph according to the piping diagram.
● Connect a stop valve upstream of each gas input to the gas chromatograph.
Gas connections on the gas chromatograph
● All gas inlets and outlets: Swagelok 1/8''.
● The gas inlets and outlets are numbered.
1
2
3
4
5
6
7
8
Figure 4-2
Gas connections
Number
Description
Comment
1
Analysis module output
Heated
2
Analysis module output
Heated
3
Sample
Heated
4
Sample
Heated
5
Carrier gas input
Not heated
6
Output from pneumatic connectors
Not heated
7
Discharge of control medium for solenoid valves and EPC
flow inserts
Not heated
8
Reserved
Unheated, closed