Description
3.6 Functions
MicroSAM
26
Operating Instructions, 06/2012, C79000-G5376-C560-07
3.6.1
Gas flow charts
As of AS09, the MicroSAM gas chromatograph is manufactured as two versions: Live and
eLive.
Solenoid valves control the respective functions in the multi-membrane valve.
Gas flow chart for analysis modules A01 - A11 (live injection)
&DUULHUJDV
(3&
(3&
7&'%
7&'%
7&'$
7&'$
(3&
6DPSOH
LQOHW
6DPSOH
RXWOHW
/LYH
,QMHFWLRQ
%DFNIOXVKLQJ
VDPSOHIORZ
/LYH
FROXPQ
VZLWFKLQJ
7&'%
7&'$
A01 ... A03
7&'$
7&'%
&ROXPQ
D
&ROXPQ
E
&ROXPQ
7&'$
A04 ... A11
7&'%
&ROXPQ
D
&ROXPQ
E
&ROXPQ
09
09
09
09
09
09
6DPSOH
ORRS
Solenoid valve 1 (MV1)
Membrane valve function 1: "Bring sample loop to pressure of EPC 1, and purge"
Solenoid valve 2 (MV2)
Membrane valve function 2: "Purge sample loop with sample"
Solenoid valve 3 (MV3)
Membrane valve function 3: "Empty sample loop completely"
Solenoid valve 4 (MV4)
Membrane valve function 4: "Inject"
EPC
Electronic pressure control
TCD
Thermal conductivity detector
Figure 3-5
Analysis module of type A with live injection
Summary of Contents for MicroSAM
Page 1: ...Process Gas Chromatography Process Gas Chromatograph MicroSAM Operating Instructions 03 2012 ...
Page 2: ......
Page 8: ...Table of contents MicroSAM 6 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 10: ...Purpose of this documentation MicroSAM 8 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 106: ...Examples MicroSAM 104 Operating Instructions 06 2012 C79000 G5376 C560 07 ...
Page 114: ...Dimension drawings MicroSAM 112 Operating Instructions 06 2012 C79000 G5376 C560 07 ...