MEMOSKOP 2/3/50/100/SUB
RX57-029.061.01
Page 2 of 2
Siemens AG
Rev. 03
10.97
TD SD 34
Medical Engineering
4 - 2
Calling up the service menu
UROSKOP D1/D2
4
•
Switch the system "ON"
•
Wait for the MEMOSKOP to boot.
Activate set-up mode
4
•
Press the following keys on the keyboard in sequence:
- <NEXT PAT>
- <@>
- <%>
- <CTRL> + <S>
After activating set-up mode, a sequence of numbers is displayed in the lower image
region of the monitor.
•
To call up the test images, overwrite this sequence of numbers.
•
Enter the number sequence without any blanks.
Depending on the test image, reconstruction may take up to 30 seconds.
Test images:
- SMPTE-similar test image
0000 8808 0A0303C F00
- Gray wedge horizontal
0000 8808 0A03021 C00
- Black/white step
0000 8808 0A03023 C00
- Gray scale 0%-100%(linear)
0000 8808 0A03025 C00
- Gray scale 0%-100% (exponential)
0000 8808 0A03026 C00
- Black 0%
0000 8808 0A03027 C00
- Vertical white bar 60%
0000 8808 0A03028 C00
•
Press keys <CTRL> + <E>.
Exit set-up mode
4
•
Switch the system "OFF" and back "ON" again.
The memory returns to normal operating mode.
<x> means: press key x.
<CTRL> + <E> means: press and hold <CTRL> and simulta-
neously press key <E>.