Important Notes
9
Page 9 of 36
Medical Solutions
Protective measures
0
System power supply
0
It is very important that any intervention in the equipment starts with disconnecting it from
the power supply with the main circuit-breaker. Before removing or inserting any of the
printed circuit boards, switch off the equipment.
WARNING
If the system is only switched off at the control panel, the line volt-
age will still be present at the generator line connection (see wiring
diagram SPB7-250.844.01...).
A life-threatening hazard of electric shock exists.
¹
Disconnect the mains cable and comply with the infor-
mation on this page.
WARNING
After shut-down of the system, there may still be 380 V DC present
on the intermediate circuit.
A life-threatening hazard of electric shock exists.
¹
Disconnect the mains cable and comply with the infor-
mation on this page.
PC boards
0
CAUTION
The p.c. boards contain electrostatic highly sensitive components.
If disregarded, the components could be damaged.
¹
Use ESD equipment, ground prior to making contact and
place the components on a conductive surface.
Metal curtain
0
WARNING
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
¹
Apply the protective strips as mentioned in section
(Link
chapter)
after removing the covers from the stand.
Remove the protective strips only when the covers are to
be mounted or when vertical adjustment of the swivel
arm system is necessary.