Final procedures
89
Page 89 of 116
Medical Solutions
11-
10Final procedures
Disabling the film exposure
0
The film exposure switch has to be
disabled
.
1. Connect the service PC to the MAMMOMAT stand and start the service software.
2. Go to:
Configuration -> Miscellaneous -> Allow film exposure
and make sure that the
Switch
is set to
OFF.
Fig. 60:
Set Allow film exposure to OFF