Mechanical covers
43
Page 43 of 116
Medical Solutions
Adapting OPCOMP
If the factory-set OPCOMP is not to the customer’s satisfaction, an adjustment can be car-
ried out with the tautness factors in the
Optimized compression dr
table, see Service
Instructions (
Software, SPB7-250.816.03...
). A lower tautness value will increase the
compression force and vice versa.
•
Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this
setting for at least two weeks before further changes are made.
NOTE
OPCOMP has been thoroughly tested by Siemens. If any factors
are changed, OPCOMP will no longer be an optimization of com-
pression force and image quality according to Siemens’ clinical
test.
Checking the exposure blocking
0
Fig. 26:
Swivel arm system_2
Check that the exposure release is blocked during the following conditions:
•
No patient is registered.
•
No compression plate is mounted.
•
The wing is not in position.