Protective measures
0
System power supply
0
Before you begin working on equipment, it is very important that you disconnect it from
the power supply at the main circuit breaker. Before removing or inserting any of the
printed circuit boards, switch off the equipment.
WARNING
If the system is only switched off at the control panel, the line volt-
age will still be present at the generator line connection (see wiring
diagram).
A life-threatening hazard of electric shock exists.
¹
Disconnect the mains cable and follow the instructions
presented here.
WARNING
After the system shuts down, there may still be 380 V DC present
on the intermediate circuit.
¹
A life-threatening hazard of electric shock exists.
PC boards
0
CAUTION
The printed circuit boards contain highly electrostatically-sensi-
tive components.
If disregarded, the components could be damaged.
¹
Use ESD equipment, ground prior to making contact, and
place the components on a conductive surface.
Metal curtain
0
WARNING
The edges of the metal curtain on the stand are very sharp.
They may cause severe injury.
¹
These strips must always be applied onto the edges of
the metal curtain during service and maintenance work.
Apply the protective strips after removing the covers from the stand. Remove the protec-
tive strips only when the covers are to be mounted or when vertical adjustment of the
swivel arm system is necessary.