Start-up Protocol
MAMMOMAT Novation DR
SPB7-250.815.06.02.02
Siemens AG
12.05
CS SD 24
Page 104 of 116
Medical Solutions
Tab. 21
AEC Dose settings
OPDOSE settings
0
Results from the chapter "Checking OPDOSE" on
.
Tab. 22
OPDOSE Settings
Check of grid lines
0
Results from the chapter "Checking Grid Lines" on
.
Tab. 23
Measuring Grid Lines
AEC
kV
Anode /
Filter
Dose value
Pixel value
H
28 kV
Mo/Mo
D
H
Mo/Rh
-
D
-
H
W/Rh
-
D
-
Program
button
Thickness
kV
Anode/filter
combination
AEC
1
2
3
4
Exposure
Exp Mode
mAs
PMMA/
mm
Tube
Angle
visible
grid
lines
(yes/no)
X
Y
Mean pixel
value differ-
ence [%]
1
mAs
32
30
0
o
2
mAs
56
40
0
o
3
mAs
200
60
0
o
4
AEC "H"
-
40
0
o
5
AEC "H"
-
40
+ 60
o
6
AEC "H"
-
40
- 60
o