10 - 1
Siemens AG
SPB7-250.815.02
Page 1 of 4
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
Final procedures
10
Settings date and time
10
MAMMOMAT Stand
Prerequisite
•
The correct date and time must be set in the service PC. If not, set the date at C:\> by
typing
date <
ENTER
>
and the time at C:\> by typing
time <
ENTER
>.
Test performance
1.
Start the Service Software for the MAMMOMAT Stand on the Service PC.
2.
In the
Mainmenu
select
Configuration
Clock.
3.
Press
<F2>
.
With
<F2>
the date and time set in the service PC are taken over by the
MAMMOMAT stand.
BRICK
10
Setting the date and time on the BRICK.
1.
Go into the service software and select Utilities -> Escape to OS and enter the
following command in the command field.
start telnet brick
2.
A username and password have to be entered.
Username:
root
Password:
******
Fig. 1