
Mechanical Checks
5 - 7
Siemens AG
SPB7-250.815.02
Page 7 of 12
MAMMOMAT
Novation
DR
Medical Solutions
Rev. 02
09.04
CS SD 24
System
The movement shall stop automatically when either the end position is reached or
when the button is released.
Emergency stop
5
1.
Press the emergency stop button (4/Fig. 6).
The button will latch and all motorized movements of the equipment will be
blocked ("Er 813" shall appear). To reset the emergency stop button, turn it
clockwise.
All eight vertical adjustment buttons (four on either side 3/Fig. 6)
must be checked. The switching off in upper and lower end posi-
tion is effected by switches S881 and S884 respectively.
NOTE