Important Notes
Protective measures
2
System Power Supply
2
It is very important that any intervention in the equipment will start with disconnecting it
from the power supply with the main circuit-breaker. Before removing or inserting any of
the printed circuit boards, switch off the equipment.
PC Boards
2
Metal Curtain
2
If the system is only switched off at the control panel, the line
voltage will still be present at the generator line connection (see
wiring diagram).
A life-threatening hazard of electric shock exists.
After shut-down of the system, there may still be 380 V DC
present on the intermediate circuit.
A life-threatening hazard of electric shock exists.
The PC boards contain electrostatic highly sensitive components.
If disregarded, the components could be damaged.
Use ESD equipment, ground prior to making contact and place
the components on a conductive surface.
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
These strips must always be applied onto the edges of the metal
curtain during service and maintenance work.