18 - 1
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 1 of 10
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Final procedures
18
Service PC and measuring instruments
18
Recording the programmed values
18
Save the programmed values at menu command "SAve config.file".
Main menu:
1.
Select Configuration <
↑
> <
↓
> < ENTER>.
2.
Select Save config.file <
↑
> <
↓
> < ENTER>.
3.
Press <F2>.
Save data with <F2>.
Reading the exposure counter and deleting the error memory
18
Read the exposure counter and delete the error memory after putting the stand into ser-
vice.
A/ Reading the exposure counter
Main menu:
1.
Select Service <
↑
> <
↓
> < ENTER>.
2.
Select Show exposure counter <
↑
> <
↓
> < ENTER>.
3.
Enter the current counter reading in the certificate.
B/ Deleting the error memory
Main menu:
1.
Select Service <
↑
> <
↓
> < ENTER>.
2.
Select Delete error buffer <
↑
> <
↓
> < ENTER>.
3.
Delete the error memory with <Y>.
Saving the configuration file
18
Main menu:
1.
Select Backup <
↑
> <
↓
> < ENTER>.
2.
Select Copy installation area to floppy <
↑
> <
↓
> < ENTER>.
3.
Select All <
↑
> <
↓
> < ENTER>.
4.
Leave the program with <F10>.
No write protection on the diskette! (Only if started from dis-
kette).
NOTICE