Checking the compression device and OPCOMP
15 - 3
Siemens-Elema AB
Register 3
SP B7-230.033.07
Page 3 of 4
Mammomat 1000/3000
Solna, Sweden
Rev. 03
11.99
SPS-UD
Installation and Start-Up
Checking the OPCOMP
15
1.
Decompress.
2.
Set the compression force to 20 kg by using the presetting knob (5/Fig. 1) and
compress.
The compression movement must be interrupted by OPCOMP before 20 kg is
reached, i.e. at 5-15 kg depending on the material (4/Fig. 1) used.
Adapting OPCOMP
15
If the factory-set OPCOMP is not to the customer’s satisfaction, an adjustment can be
done with the tautness factors in the "Best compression table", see Service Instructions. A
lower tautness value will increase the compression force and vice versa.
•
Increase/decrease the tautness factor by e.g. 10 the first time. Let the customer use this
setting for at least two weeks, before further changes are made.
Opcomp has been thoroughly tested by Siemens. If any factors
are changed, OPCOMP will no longer be an optimization of com-
pression force and image quality according to Siemens’ clinical
test.
NOTICE