Mammomat 1000/3000
Register 3
SP B7-230.033.07
Page 2 of 4
Siemens-Elema AB
Installation and Start-Up
Rev. 03
11.99
SPS-UD
Solna, Sweden
19 - 2
Bolting the stand/generator to the floor
Installation with separate generator
19
The separate generator must be bolted to the floor and/or be secured to the wall by
means of the enclosed wall brackets. For bolting the separate generator to the floor,
the same procedure as for generator with integrated control panel and radiation shield
applies, see page 19 - 1.
For attaching the wall brackets to the generator , the uppermost screw hole for the lead-
glass stanchions shall be used. Ensure that the wall brackets are secured to the wall with
adequate safety margins. Dowels and screws are not supplied. Please obtain these
locally and ensure that they are suitable for the material used in the wall.
Fig. 3
Base plate and wall brackets for separate generator
SIEMENS
Base plate for separate generator
Predrilled
fixing holes
Generator secured with wall brackets
∅
8 mm
M
A
M
0020
9
M
A
M
0021