Operator Responsibilities for Maintenance
13
© Siemens, 2009
M7-000.664.01.04.02
Page 13 of 14
MAGNETOM
03.12
CS SD MR
Tim dockable table
Preventive measure to
avoid wear and tear
Check the docking mechanism
Check the hydraulic system
Check/refill hydraulic oil
Check/clean connectors of the docking
station
Lubricate RF connectors
Annually
RF connectors
Preventive measure to
avoid wear and tear
Check/clean connectors
Lubricate RF connectors
Annually
Comfort kit (option)
Preventive measure to
avoid overheating
Replace air filter
Every
2 years
Software Preventive
measure
against data loss in the
case of hard disk failure
Save dynamic data
Clean up directories
Every
3 months
1. According to manufacturer recommendations
2. According to manufacturer recommendations
3. Zero helium boil off rate for typical clinical use, depending on sequence and operating time with running helium compressor.
The system needs to be serviced at regular intervals. Undisturbed magnet cooling for 24 hours and 7 days a week.
4. If the system is not connected to Siemens Remote Service (SRS) the replacement has to be performed according to manu-
facturer recommendations. The replacement is performed “on demand” when the system is connected to Siemens Remote
Service (SRS).
Object or function
Reason
What is checked:
Interval
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...