If the protocol is configured for repeated measurements, the next
measurement can be started with the footswitch after completing the
preceding measurement. For more information regarding the
configuration of protocols, see:
Software operator manual
1
Push the hose plug into the retaining rings of the pushbutton unit.
2
Load a suitable protocol and start it at the syngo Acq WP or the In-
Room syngo Acq WP.
The MR system is waiting for the manual start of the protocol.
3
Press the Start footswitch to begin the measurement.
As an alternative, you can start and end the measurement via the
In-Room syngo Acq WP or the syngo Acq WP.
4
Press the Stop footswitch to end the measurement.
Other components and accessories
Gradient supervision
To prevent damage to the MR system by a malfunction of the
gradient system, a specially designed supervision is installed at your
MAGNETOM system.
This supervision monitors that cables, connections, or other
components of the gradient system do not show excessive heating. In
case of a malfunction, the measurement is stopped and an alarm
message is issued.
After 1 minute the system will be automatically switched to
Standby.
3.10
3.10.1
3
MR system components
80
MR System | Operator Manual
Print No. M6-04401.621.02.02.02