WARNING
When using the DIXON method, water and fat swaps might
occur!
Incorrect diagnosis
◆
Diagnosis should be confirmed by a second contrast and/or a
different orientation.
CAUTION
Poor image quality! Wrong image position!
Incorrect diagnosis
◆
Prior to the examination, inform patients about movements
and their negative effects on the measurement.
◆
Ensure that the patient does not move during the
measurement.
◆
Monitor the patient during the MR examination.
CAUTION
Imprecise localization due to patient movement during
functional MR imaging (fMRI)!
Incorrect stereotactic planning and injury of the patient
Incorrect assignment of active brain areas
◆
Prior to the examination, inform patients about movements
and their negative effects on the measurement.
◆
Use Siemens scan protocols with motion correction.
◆
Monitor the patient to ensure that the task is performed
correctly.
DIXON swapping
Patient-related artifacts/
imaging errors
Safety
2
MAGNETOM Skyra
fit
| syngo MR E11
37
Print No. M6-04401.621.02.02.02