LITHOSTAR Multiline
RXL2-120.812.01
Page 14 of 24
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
4 - 14
Connections with POLYDOROS SX 65
Connections/cable locations in the generator cabinet
4
•
Route the cables in the generator cabinet as shown in Fig. 8 above, and connect them in
accordance with Table 9:
•
Install the yellow cable harness in the generator cabinet below M16. Connect the
shielding of the zipper hose to the
bolt in the M16 (
/Fig. 8).
Strain-relieve the zipper hose.
Rotating anode cable connection
•
Guide the rotating anode cable (internally required length approx. 1 m) up along the right
side in the frame of the generator cabinet to the anode starter (
/Fig. 8).
•
Connect the braided shielding (external shielding) of the end of the lead to the right side
of the anode starter under one of the clamps (
/Fig. or 1/Fig. 9).
•
Connect the rotating anode cable to the K3 breaker of the anode starter (
/Fig. 8):
On breaker, contact:
K31/R2
K31/R4
K32/R2
Anode cable leads:
I / 1
0
II / 2
D
F
F
F