LITHOSTAR Multiline
RXL2-120.812.01
Page 8 of 20
Siemens AG
System
Rev. 02
10.04
CS PS 24
Medical Solutions
3 - 8
Connections with POLYDOROS LX 30
Fig. 5
Connections and cable locations in the
A12 electronics cabinet
and in the control console
3
(cable harness, green)
•
Route the cables in the A12 electronics cabinet,
as shown in Fig. 5 and connect them as shown in
Table 5, 6 and 7.
VIDEOMED connection:
•
Remove the XCS cables X11 and X12 from the D5
board.
•
Remove the EMC shielding on the Videomed.
•
Route the camera connector D8.X1similar to the video
cable underneath the rack and connect it to the
backplane as shown in Table 5.
Remove board D1 for this purpose.
•
Attach the cable with a cable clip to the rack above the
shield.
•
Attach the EMC shielding on the Videomed.
•
Connect the XCS cables X11 and X12 on the D5 board.
8
8
7
6
9
4
4
5
10
A12
Mounting rail for cable harness
Cable harness, green, from the control console
Cable harness, green, from the unit (K3, K4....)
Cable harness, green, from the unit (K8, K9....)
3rd monitor on the monitor trolley (option)
Video cable 3rd monitor
VIDEOMED
-DIM-
Power Supply
M16
System
Controller
N11
Image Storage