System Installation
2 - 9
Siemens AG
RXL2-120.812.01
Page 9 of 14
LITHOSTAR Multiline
Medical Solutions
Rev. 02
10.04
CS PS 24
System
Installing the Components for the System Controller
2
Setting Up the Console Table
2
The cable outlet (A/Fig. 16) is set up at the factory for downward cable feed-through.
However, it is possible to route the cables to the back by punching out one of the perfo-
rated openings.
•
Route the cable from the control console table to the control cabinet.
•
If the Diamentor option is present, route the Diamentor cable from the electronics cabinet
to the console table.
Positioning/Connecting the Monitors
2
•
Place the monitors on the control console table as shown in Fig. 15 and connect them
per their designations.
Positioning the Keyboard and the PC Mouse
2
•
Place the keyboard and the PC mouse on the control console table.
•
Connect the cable to the keyboard.
Connecting the Footswitch, Tableside Control Console,
Shock Wave Trigger Mechanism
2
•
Connect the tableside control without the holder (optional) (B/Fig. 16) and the (C/Fig. 16)
shock wave release control to the console table.
•
Connect the footswitch for fluoroscopy (optional) either to the console table or to the unit:
- Right back panel of the unit,
- Under the foot support at the control console table.
•
Connect the tableside control with the holder to the unit.
•
Connect the unit footswitch (optional) on the right in back of the unit.
Fig. 15
Fig. 16
B
C
A
PC-
Monitor
Monitor I + II
Multispot
or DICOM
Connect
(Options)
Operating
system
unit
DICOM
Monitor
(Option)
p