LITHOSTAR Multiline
RXL2-120.815.01
Page 2 of 18
Siemens AG
System manual
Rev. 02
02.05
CS PS 24
Medical Solutions
6 - 2
Checking the Dose Rate
Flowchart
6
The IQC program and IQC data disks are located in the in the sys-
tem binder. Instructions can be taken from the IQC program
instructions RXD0-000.073.... The installation text appears in Ger-
man, however other languages can be set.
The following flowchart displays only the basic steps of the IQ
tests.
NOTE
Is the IQC Program
installed on the service PC?
no
yes
no
yes
Loaded
protocol
"open"
Start IQC Program
Load data from the system disk with
Copy data
→
Load data from file...
→
ALL
Close protocol with
Manager
→
close
protocol
Open start-up protocol with
Manager
→
Open new protocol
and confirm with
Yes
Select IQ start-up and confirm query
"DHHS"
or
"NO"
(Menu option
§ 16 X-Ray Ordinance
is presently not used)
continue with ESC
Refer to instructions IQC-
Program RXD0-000.073
...
Install the program on the
DOS level.