6 - 1
Siemens AG
RXL2-120.815.01
Page 1 of 18
LITHOSTAR Multiline
Medical Solutions
Rev. 02
02.05
CS PS 24
System manual
Checking the Dose Rate
6
General Remarks
6
The following image quality tests for the Test Protocol ensure that the system is function-
ing properly and that the factory settings have not changed.
•
Perform the IQ test using an IQC program.
•
The IQC program must already be installed on the service PC or must be installed
according to RXD0-000.73.01...
•
The IQC program disk and the IQC data disk containing the corresponding IQC data
(RXD0-000.037.04.., item no. 37 67 410 X1963) are located in register 9 of the system
binder.
Test Videomed DIM
6
After an adequate warm-up time for Videomed DI:
Select the service software VIDEOMED DIM.
Use the IQC floppy disk, RXD0-000.037.04... to record the data.
In Menu
→
Adjustment
select submenu
Gamma calibration
and
Ramp calibration
and perform the steps indicated.
NOTE