Parameter reference
8.7 Maintenance and Diagnostics (3.)
HydroRanger 200 HMI
262
Operating Instructions, 06/2018, A5E36281317-AC
TS-3 temperature minimum (3.2.10.5.)
Shows the lowest temperature encountered, as measured by the TS-3 Temperature Sensor
(if applicable).
Index
Global
Values
Range: - 50 ... 150 °C (Read only)
Preset: 150 °C
Related
TS-3 temperature maximum (3.2.10.4.) (Page 261)
To reset the log after an open circuit on the transducer wiring, enter the value 150.
8.7.2.11
Echo quality (3.2.11.)
Transducer selector (3.2.11.1.)
Sets the transducer index for all parameters applicable to this sub-menu.
Long confidence (3.2.11.2.)
Displays the Long Echo Confidence of the measurement echo from the last shot.
Index
Transducer
Values
Range: 0 ... 99
Related
•
Long echo threshold (2.11.2.3.) (Page 223)
•
TVT type (2.11.3.6.) (Page 230)
Use this feature to monitor the effect of transducer aiming, location, and mechanical
transducer/mounting isolation.
Short confidence (3.2.11.3.)
Displays the Short Echo Confidence of the measurement echo from the last shot.
Index
Transducer
Values
Range: 0 ... 99
Related
•
Long echo threshold (2.11.2.3.) (Page 223)
•
TVT type (2.11.3.6.) (Page 230)
Use this feature to monitor the effect of transducer aiming, location, and mechanical
transducer/mounting isolation.