Page 1—4, Section 1: Introduction
Theory of Operation
The Hematek Slide Stainer is designed to produce stained slides of consistent quality in a
continuous process. This is accomplished by having a fixed length of time in each of the
three phases of stain, buffer, and rinse, as well as a predetermined ratio of stain-to-buffer
volumes in the buffer phase.
A properly stained slide is the result of an interactive process involving the pump
volumes, the mixing process, and the stain-to-buffer ratio. A minimum ratio of 1:2 is
recommended; however, if the platen is filling properly, good mixing is occurring, and the
stained slides are acceptable under the microscope, the stain-to-buffer ratio should not be
adjusted.
Figure 1-1. Hematek Slide Stainer
Summary of Contents for Hematek
Page 1: ......
Page 6: ...Contents ...
Page 8: ...Page 1 2 Section 1 Introduction ...
Page 22: ...Page 2 2 Section 2 Installation ...
Page 32: ...Page 3 2 Section 3 Operating Instructions ...
Page 38: ...Page 3 8 Section 3 Operating Instructions ...
Page 40: ...Page 4 2 Section 4 Specimens ...
Page 44: ...Page 4 6 Section 4 Specimens ...
Page 46: ...Page 5 2 Section 5 Maintenance ...
Page 54: ...Page 5 10 Section 5 Maintenance ...
Page 56: ...Page 6 2 Section 6 Minor Replacements and Adjustments ...
Page 76: ...Page 7 2 Section 7 Troubleshooting ...
Page 82: ...Page 8 2 Section 8 Service Supplies and Replacement Parts ...
Page 88: ...Page 8 8 Section 8 Service Supplies and Replacement Parts ...
Page 94: ...Page B 2 Appendix B Hematek Preservice Checklist ...
Page 98: ...Index 4 ...