Functions
7
7.1
UniMass Table
Some SITRANS F 1010 family flow meters provide a feature called UniMass. These flow
meters support a UniMass Table which associates the metered liquid’s current temperature
and sonic velocity (Vs) with its specific gravity and viscosity. These variables are then used
to provide mass metering and real-time flow profile compensation based on Reynolds
numbers.
Note
For hydrocarbon liquids requiring standard volume compensation, an FUH1010 Standard
Volume flow meter should be specified.
The table scanning routine can function with a minimum of two completely defined data
points (e.g. Temperature and Specific Gravity). UniMass will not project viscosity and
specific gravity measurements beyond the entered limits of the table but rather keep the
highest or lowest table entry as its output.
Building the UniMass Table
The table is built by inputting a series of data points—two inputs and two outputs each:
●
Input data consists of sonic velocity (Vs) and temperature.
●
Output data consists of specific gravity (S.G.) and viscosity (cS) at the specified Vs and
temperature.
UniMass Table Operation
The UniMass Table is designed to operate with individual families of liquids.
Although the UniMass Table is designed for variable temperature and sonic velocity,
Constant Vs or Constant Temp table editing capabilities are provided as well.
Application Data Menu Definitions for UniMass Table
The UniMass Table resides within the [Application Data] menu. Refer to Programming the
Transmitter (Page 49) for Installation Menu details.
[UniMass Table] - From the [Application Data] menu, press the <Right Arrow> to access the
UniMass data point editor.
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
93
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...