Setup
3.1 Sensor settings
FST020 Function manual
34
Function Manual, 09/2017, A5E42949526-AA
3.1.9.3
Calibration point
Description
Calibration point that is to be transferred to the multipoint calibration table.
Setting 0.0 m
3
/s to 100000.0 m
3
/s
Default 0.0 m
3
/s
Related
Path
Setup
→
Sensor settings
→
Calibration point
3.1.9.4
Multipoint calibration table
3.1.9.4.1 Calibration points 1 - 20
Description
Calibration points 1 through 20 for unidirectional flow or calibration points 1 through 20 for
bidirectional positive flow. If 0.0 the calibration point is not used.
Access level: Expert user only.
Note
Only available for PID_ProtectCalibrationTable == 1 OR PID_EnableUserCalibrationTable ==
1
Setting 0.0 m
3
/s to 100000.0 m
3
/s
Default 0.0 m
3
/s
Related
Path
Setup
→
Sensor settings
→
Multipoint calibration
→
Multipoint calibration table
→
Calibration
points 1 - 20
3.1.9.4.2 Calibration value
Description
Calibration correction factor that is associated to Calibration points 1 through 20.
Access level: Expert user only.
Setting 0.5 to 2.0
Default 1.0