Setup
3.1 Sensor settings
FST020 Function manual
26
Function Manual, 09/2017, A5E42949526-AA
3.1.6.2
Clamp-on sensor size
Description
Installed sensor size. Read only. Can only be changed via the Sensor Wizard. Indicates the
Sensor model chosen from the sensor setup wizard.
3.1.6.3
Nominal sensor frequency
Description
Nominal frequencies for the sensors
Setting 80000.0 m/s to 4000000.0 m/s
Default 1100000.0 m/s
Related
Path
Setup
→
Sensor settings
→
Sensor selection
→
Nominal sensor frequency
3.1.6.4
Sensor crystal projection
Description
Length of the crystal projection onto the emitting surface of the sensor (clamp-on). Access
level: Expert user only.
Setting 0.0 m to 0.3 m
Default 0.03409 m
Related
Path
Setup
→
Sensor settings
→
Sensor settings
→
Sensor crystal projection
3.1.6.5
Sensor phase velocity
Description
The velocity at which the wave fronts traverse the interface between the wedge and pipe.
Access level: Expert user only.
Setting 1000.0 m/s to 20000.0 m/s
Default 3695.5 m/s
Related
Path
Setup
→
Sensor settings
→
Sensor settings
→
Sensor phase velocity
Summary of Contents for FST020