Description
3.4 Method of operation
FIDAMAT 6
26
Operating Instructions, 01/2019, A5E00222135-04
3.4
Method of operation
①
Flame
⑤
Sample gas
②
Measuring electrode
⑥
Combustion air
③
Amplifier
⑦
Quartz nozzle
④
H
2
Figure 3-2
Method of operation of flame ionization detector
Note
Differences in the method of operation for the device versions
Unless stated otherwise, the description relates to both device versions of the FIDAMAT 6.
The measuring principle is based on flame ionization detection (FID) with a detection limit of
approximately 0.1 ppm. During the combustion of hydrocarbons in a hydrogen flame, ions
are formed that are in turn converted into a current flow by the application of an electric field.
The resulting amperage is a measure of the number of carbon atoms.
The FIDAMAT 6 measures specific to the substance class, not component-specific. It
measures the total of all organically bound carbons in a sample gas, however with different
weighting of the hydrocarbon molecules.
Summary of Contents for FIDAMAT 6
Page 1: ...FIDAMAT 6 ...
Page 2: ......
Page 10: ...Table of contents FIDAMAT 6 8 Operating Instructions 01 2019 A5E00222135 04 ...
Page 38: ...Description 3 6 Communication FIDAMAT 6 36 Operating Instructions 01 2019 A5E00222135 04 ...
Page 76: ...Operation 7 6 Calibration FIDAMAT 6 74 Operating Instructions 01 2019 A5E00222135 04 ...
Page 134: ...Functions 8 6 Configuration FIDAMAT 6 132 Operating Instructions 01 2019 A5E00222135 04 ...
Page 176: ...Dimensional drawings FIDAMAT 6 174 Operating Instructions 01 2019 A5E00222135 04 ...
Page 186: ......
Page 198: ...Glossary FIDAMAT 6 196 Operating Instructions 01 2019 A5E00222135 04 ...
Page 201: ......
Page 202: ......