Table 7-10 Technical data - gas input conditions
Gas input conditions
Allowed sample gas pressure
approx. 1 000 hPa
Sample gas flow
18 to 60 l/h (0.3 to 1 l/min)
Sample gas temperature
0-200 °C
Sample gas moisture content
< 90 % RH (RH = relative humidity)
Table 7-11 Technical data - pressures and flow
Pressures and flow
Gases
Input pressure
hPa (rel.)
Operating
pressure
hPa (rel.)
Flow via FID
ml/min.
Flow via bypass
ml/min.
Combustion gas
2.000 – 4.000
1.000 ± 20
~ 25
-
Combustion air
2.000 – 4.000
475 ± 5
~ 320
~ 500
Sample gas
800 – 1.000
-
~ 3
~ 1.000
Zero gas
800 – 1.000
-
~ 3
~ 1.000
Test gas
800 – 1.000
-
~ 3
~ 1.000