Operation
5.11 Configuration
FIDAMAT 6-G
Operating instructions, 07/2006, A5E00478463
5-61
With this function, you switch the message for maintenance requests and faults off
individually so that there is no entry in the logbook, no message in the status line of the
measured value display and no external signal sent.
5.11.19
PROFIBUS configuration (function 90)
9O Profibus config. THC
Adress :126:
Ident number: O
Software release: 1.4.O
Figure 5-68 Configuring PROFIBUS
The function can only be called if the device contains additional PROFIBUS electronics
(optional board).
With this function, you can set the PROFIBUS station address. The address can be set from
0 to 126.
The configuration behavior of the device is configured with the
"Ident number" (softkey 2).
Parameters 0, 1 and 3 can be selected and have the following meaning:
Parameter
Meaning
0
Only the PROFILE ID number is positively acknowledged
1
Only the device-specific ID number is positively acknowledged
3
Only the PROFILE ID number for multivariable devices (complex analyzers) are
positively acknowledged.