Process value
Modbus register
Units
Density
7600
• µg/l (micrograms per liter)
• µg/m³ (micrograms per cubic meter)
• mg/l (milligrams per liter)
• g/ml (grams per milliliter)
• g/cm³ (grams per cubic centimeter)
• g/l (grams per liter)
• kg/l (kilograms per liter)
• kg/m³ (kilograms per cubic meter) (default)
• lb/in³ (pounds per cubic inch)
• lb/gal (pounds per US gallon)
• lb/ft³ (pounds per cubic foot)
• ton/yd³ (short tons per cubic yard)
• Custom
Temperature
7700
• °C (degrees Celsius) (default)
• °F (degrees Fahrenheit)
• °R (degrees Rankine)
• K (kelvins)
Mass totalizer
• g (grams)
• kg (kilograms)
• t (metric tons)
• ton (short tons)
• ton (UK) (long tons)
• oz (ounces avoirdupois)
• lb (pounds)
• Custom
8.1.2
Zero point adjustment
See also the description of the Zero point adjustment via SIMATIC PDM (Page 62).
Note
Preconditions
Before a zero point adjustment is initiated, the pipe must be flushed, filled and at an absolute
flowrate of zero preferably also at operating pressure and temperature.
Note
Change of parameters during zero point adjustment
Do not change any other parameter during the zero point adjustment procedure.
Parameter assignment FCT010
8.1 Functions
SITRANS MASS 2100 & FC300 (FCT010/DSL)
76
Operating Instructions, 06/2020, A5E39788682-AB