8. During the process a progress bar is visible.
9. At the end of the zero point adjustment the outcome is displayed as an offset and a standard
deviation.
Note
If you get an error message after the zero point adjustment, refer to Fault codes and
corrective actions (Page 96).
The system is now ready for normal operation.
6.3.8
Changing parameter settings using SIMATIC PDM
Note
Clicking on "Cancel" during an upload from device to SIMATIC PDM will result in some parameters
NOT being updated.
Many parameters are accessed via the online menus in SIMATIC PDM, see Parameters accessed
via drop-down menus (Page 66) for the others.
1. Launch SIMATIC PDM, connect to the appropriate device and upload data.
2. Adjust parameter values in the parameter value field.
3. Click on "Enter".
The status fields read "Changed".
4. Open the menu "Device".
5. Click on "Download to device".
6. In menu "File" click "Save" to save settings offline.
The status fields are cleared.
Commissioning
6.3 Remote commissioning with PDM FCT010
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
65