5. Click on "Configure selected device".
6. Go to the tab "FW update".
7. Click "Load FW file".
8. Select firmware bundle.
9. Click "Start FW update".
10.3.2
Troubleshooting sensor-related problems
Incorrect and unstable measurements, especially at low flows, are typically a result of an
unstable zero point due to:
• Incorrect installation
• Bubbles in the liquid
• Vibrations or cross talk
• Solid particles settling in the liquid
Follow the 4-step guide to troubleshooting:
Step 1
Preliminary application inspection (Page 102)
Step 2
Zero point adjustment (Page 102)
Diagnostics and troubleshooting FCT010
10.3 Operation troubleshooting
SITRANS MASS 2100 & FC300 (FCT010/DSL)
Operating Instructions, 06/2020, A5E39788682-AB
101