Operation
Setting ranges
Ventilation parameter
Setting range
Inspiration pressure P
insp
*
[mbar]
*
Only values can be set which are 2 mbar above the set PEEP value
LeakV: 5 to 40
ExpV: 5 to 50
PEEP
**
[mbar]
** An increase in PEEP value necessitates an increase in the Pinsp value when a min-
imum difference of 2 mbar has been reached
3 to 20
Inspiratory O
2
concentration FiO
2
[%]
21 to 100
Trigger sensitivity Trigger
Off/Normal/Sensitive
Frequency f [bpm]
5 to 50
Inspiration time Ti [sec]
or I:E
0.3 to 8
1:39 to 2:1
Ramp rise time Ramp [sec]
Auto/0.1 to 2.0