Operation
66
Instructions for Use Carina SW 3.n
PC-BIPAP
Pressure Control – Biphasic Positive Airway
Pressure
Pressure-controlled ventilation in conjunction with
free spontaneous breathing during the entire
breathing cycle. The mandatory portion of total
minute volume MV is adjusted with inspiration pres-
sure P
insp
, PEEP and frequency f.
Suitable for patients without spontaneous breath-
ing, to spontaneous breathing patients before extu-
bation. Withdrawal by gradually reducing the
mandatory portion of total minute volume MV.
Set PC-BIPAP with the buttons for the ventilation
parameters:
– Inspiration pressure »
P
insp
«
– Positive end expiratory pressure »
PEEP
«
– O
2
concentration »
FiO
2
«
– Sensitivity of »
Trigger
« (for synchronising
spontaneous breathing)
– Frequency »
f
«
– Inspiration time »
Ti
« or »
I:E
« (configurable, see
– Pressure rise time »
Ramp
«
z
Non-invasive ventilation with a hose system
with leak valve – using the »
Mask
« application
method, see page 75.
z
Setting alarm limits, see page 84.
036
Flow
Paw
Ti
f
T=
PEEP
Pinsp
Ramp high
Ramp low
t
t
1
Trigger window
without spontane-
ous breathing
with spontaneous
breathing
522
523
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...