Start-up
60
Instructions for Use Carina SW 3.n
Setting the O
2
supply
Carina has been set to HPO mode by the manufac-
turer.
For LPO mode, configure LPO, see page 78.
Dräger recommends the use of an external oxygen
monitor add to the patient circuit to measure the
actual oxygen concentration. Carina displays the
concentration set by the user, but does not denote
the actual concentration delivered to the patient.
Placing the oxygen sensor
The oxygen sensor for the oxygen monitor should
be placed directly at the patient connection port of
Carina, before the mask or tube.
093