58
Instructions for Use Carina SW 3.n
Start-up
Checklist
Action
Observe
Check
1
Connect all necessary equipment: hoses,
cable, filter, etc.
Equipment is connected correctly, refer to
page 31.
2
Turn the main switch on.
Carina performs a self-test, LEDs light up
and alarm tone is given.
3
Check if the used hose system (leak valve or
expiratory valve) conforms with the text in the
status line.
»
LeakV
« for leak valve or »
ExpV
« for expir-
atory valve is displayed in the status line.
4
Check if the used patient connection (tube or
mask) conforms with the text in the status line.
»
Mask
« for mask or »
Tube
« for tube appli-
cation method is displayed in the status line.
5
Connect the test lung to the hose system.
6
Make following settings:
Mode
PC-BIPAP
Settings
P
insp
15 mbar
PEEP
5 mbar
FiO
2
21 %
f
10 bpm
Ti/I:E
2 s/1:2
Ramp
0.2 s
Alarms
Paw
high 20 mbar
T
disconn
0 s
Press »
Start/ Standby
« button to start the ven-
tilation and squeeze the test lung several times
strongly during inspiration.
Each time the lung is compressed, expira-
tion should be initiated instantly.
After the second compression on P
insp
level
alarm tone starts and »
Airway pressure
high !!!
« is displayed.
7
Confirm alarm, then disconnect the test lung
from the hose system.
Alarm tone starts and »
Airway pressure
low !!!
« is displayed.
8
Reconnect test lung and hose system.
9
Change following settings:
Mode
SPN-CPAP
Settings
Apn.Vent. on
Alarm
T
apn
5 s
Steadily inflate and deflate the test lung to
simulate spontaneous breathing, then
stop simulation.
Without spontaneous breathing and after
apnoea alarm time has elapsed, alarm
»
Apnoea ventilation !!!
« is displayed.
Apnoea ventilation starts.
10
Set ventilation mode to PC-BIPAP.
11
Disconnect device from all external power sup-
plies.
Alarm tone starts and display indicates
»
External power failure !!
«.
12
Reconnect the power supplies. Disconnect the
test lung from the hose system.
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...