Instructions for Use Carina SW 3.n
51
Preparation
HPO – High Pressure Oxygen mode
For O
2
supply to Carina from the central gas supply
or from a O
2
gas cylinder.
z
Screw the O
2
compressed gas hose onto the
HPO nozzle of Carina. Supply with oxygen from
the central supply or from the O
2
gas cylinder.
LPO – Low Pressure Oxygen mode
For O
2
supply to Carina from a mobile oxygen
source, separate from a central high-pressure O
2
supply.
1
Remove the cap from the LPO nozzle for O
2
supply.
2
Connect the O
2
supply hose from the O
2
source
to the LPO nozzle.
3
Connect a flow control valve on the O
2
cylinder
with pressure gauge or a concentrator.
Observe the Instructions for Use for the O
2
components.
WARNING
O
2
compressed gas must be dust-free, oil-free
and dry and its pressure should be between
2.7 and 6 bar (270 and 600 kPa). Risk of fire.
WARNING
Only use O
2
gas supplies with 100 % O
2
. Risk
of incorrect FiO
2
at patient.
021
02
2
2
1
3
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...