Description
128
Instructions for Use Carina SW 3.n
Compensation of volume and pressure losses
Carina with hose system with leak valve compen-
sates leakages from valve and patient.
However, to compensate the volume loss due to
the compliance of the different hose systems, con-
sider the following table:
MV
lost
(minute volume lost) = VT
lost
(tidal volume
lost) x f (frequency).
Hose system
Δ
P above
PEEP
Volume loss
(VT
lost
)
MP 00 312/
21 65 643
5 mbar
4 mL
MP 00 312/
21 65 643
10 mbar
7 mL
MP 00 312/
21 65 643
15 mbar
11 mL
MP 00 313
5 mbar
4 mL
MP 00 313
10 mbar
7 mL
MP 00 313
15 mbar
11 mL
MP 00 313
20 mbar
14 mL
MP 00 313
25 mbar
17 mL