Technical Data
116
Instructions for Use Carina SW 3.n
Inspiratory tidal volume VTi,
VT
Range
0 to 4000 mL, BTPS
Resolution
1 mL
Accuracy
±10 % of the measured value (without leakages)
Waveform display
Airway pressure Pressure (t)
0 to 80 mbar
Flow (t)
–160 to +160 L/min
Bar graph display
Airway pressure Pressure
0 to 60 mbar
Monitoring
Inspiratory minute volume MV
Alarm upper alarm limit
If the upper alarm limit has been exceeded.
Setting range
2 to 40 L/min, in 0.1 L/min increments
Alarm lower alarm limit
If the lower alarm limit has been exceeded.
Setting range
0.1 to 39 L/min, in 0.1 L/min increments
Airway pressure Paw
Alarm upper alarm limit
If the "Paw high" value has been exceeded.
Setting range
10 to 55 mbar, in 1 mbar increments
Delay time T
disconn
Mask ventilation
0, 15, 30, 60, 90, 120 seconds
Tube ventilation
0, 15, 30, 60 seconds
Apnoea alarm time T
apn
Alarm
If no respiratory activity is detected.
Setting range
5 to 60 seconds, adjustable in 1 second increments.
Breathing frequency f high
Alarm
If the alarm limit has been exceeded for 1 minute.
Setting range
10 to 50 bpm
Summary of Contents for Drager medical Carina
Page 4: ...4 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 24: ...24 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 30: ...30 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 54: ...54 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 102: ...102 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 124: ...124 Instructions for Use Carina SW 3 n This page intentionally left blank ...
Page 143: ...Instructions for Use Carina SW 3 n 143 This page intentionally left blank ...