DCA Vantage Operator’s Guide
201
Computer Interface (LIS) C
o
mputer Interface
(LIS)
Architecture
The basic architecture is point-to-point communications between LIS and the
DCA Vantage system.
Hardware Link
The hardware link consists of serial asynchronous (RS-232) interface.
Communication is normally full duplex at one of the following user-selected
baud rates:
•
19200
•
9600
•
4800
•
2400
The system interface port is equipped with a 9-pin, D-type, male connector.
The systems are considered to be Data terminal Equipment (DTE). The signal
lines that are supported are listed below.
Character Format
You can select the following character format options when you select the LIS
communication parameters in setup at the DCA Vantage system.
The stop bits are set to 1.
Signal Line
Pin Number Type
Data Carrier Detect (not used)
DCD
Pin 1
Control
Received Data
RD
Pin 2
Data
Transmitted Data
TD
Pin 3
Data
Data Terminal Ready (not used)
DTR
Pin 4
Control
Signal Ground
G
Pin 5
Ground
Data Set Ready (not used)
DSR
Pin 6
Control
Request to Send (not used)
RTS
Pin 7
Control
Clear to Send (not used)
CTS
Pin 8
Control
Ring Indicator (not used)
RI
Pin 9
Control
Option
Selection
data bits
7 bits, 8 bits
parity
odd, even, none
Summary of Contents for DCA Vantage
Page 10: ...10 DCA Vantage Operator s Guide ...
Page 24: ...24 DCA Vantage Operator s Guide ...
Page 40: ...40 DCA Vantage Operator s Guide ...
Page 104: ...104 DCA Vantage Operator s Guide Maintenance Figure 36 Filter Location 1 Filter Cover ...
Page 120: ...120 DCA Vantage Operator s Guide ...
Page 150: ...150 DCA Vantage Operator s Guide File Management ...
Page 166: ...166 DCA Vantage Operator s Guide ...
Page 192: ...192 DCA Vantage Operator s Guide ...