CR-IR347
Service Manual
PM - 10
009-058-00
10.20.2000 FM2732 (2)
PM - 10
2.
Checks Before Maintenance
0.5Y
At intervals of half year
→
Check.
2.1
Checking the Error Log
(1) Start the M-Utility.
(2) Open the error log.
REFERENCE
Below is shown an example where errors and warning messages are displayed. To
check the number of occurrences for each error code, select “2. SUMMARY”, instead of
“1. ALL”, at step #3.
0. QUIT
1. ERROR LOG UTILITY
2. CONFIGURATION SETTING
3. TEST MODE
4. ELECTRICAL UTILITY
5. SCANNER UTILITY
6. MECHANICAL UTILITY
7. FILE UTILITY
8. BACKUP MEMORY
9. HV OFF
10. MENU SETTING
11.SYSTEM UTILIT
Y
> 1
0. QUIT
1. LIST
2. CLEAR
3. SAVE TO FD
4. SAVE TO HD
ELU > 1
0. QUIT
1. ALL
2. SUMMARY
ELU : LI > 1
0. QUIT
1. ERROR
2. WARNING
3. BOTH
ELU : LI : ALL > 3
FR7H6101.EPS
#1 [Input/ENT]
#2 [Input/ENT]
#3 [Input/ENT]
#4 [Input/ENT]
Summary of Contents for CR-IR 347
Page 2: ...CR IR347 CR IR347P Service Manual Preventive Maintenance PM ...
Page 22: ...CR IR347 Service Manual PM 15 3 009 058 03 08 30 2002 FM3476 PM 15 3 BLANK PAGE ...
Page 23: ...CR IR347 Service Manual PM 15 4 009 058 03 08 30 2002 FM3476 PM 15 4 BLANK PAGE ...
Page 29: ...CR IR347 Service Manual PM 21 009 058 00 10 20 2000 FM2732 2 PM 21 BLANK PAGE ...
Page 59: ...CR IR347 Service Manual PM 51 009 058 00 10 20 2000 FM2732 2 PM 51 BLANK PAGE ...
Page 73: ...CR IR347 Service Manual PM 65 009 058 00 10 20 2000 FM2732 2 PM 65 BLANK PAGE ...
Page 83: ...CR IR347 Service Manual PM 75 009 058 00 10 20 2000 FM2732 2 PM 75 BLANK PAGE ...
Page 89: ...CR IR347 Service Manual PM 81 009 058 00 10 20 2000 FM2732 2 PM 81 BLANK PAGE ...
Page 110: ...CR IR347 Service Manual PM 102 009 058 00 10 20 2000 FM2732 2 PM 102 BLANK PAGE ...
Page 119: ...CR IR347 Service Manual PM 111 009 058 00 10 20 2000 FM2732 2 PM 111 BLANK PAGE ...